- Σταθερή παροχή 180 T/h ή 300 T/h, ιδανικό για κλινικές
- Κλειστά ολοκληρωμένα φωτοβολταϊκά συστήματα, χωρίς μηχανικά μέρη, χωρίς συντήρηση
- Ο ελάχιστος όγκος αντίδρασης είναι 150 ul, εξοικονόμηση κόστους
- Διακριτή, τυχαία πρόσβαση, αυτόματη διαδικασία
- Τυχαία επιλογή μεθόδου, ο πραγματικός αυτόματος χημικός αναλυτής διάκρισης
ΕΛΕΓΧΟΙ
- -ALT, AST, TP, ALB, TBIL, DBIL, ALP, r-GT, CHE, TBA, 5-NT
-UREA, CR, UA, TCO2
-CHOL, TG, HDL, LDL, ApoA, ApoB, LPa
-AMY, LIP
-GLU, GSP
-CK, CK-MB, a-HBDH, LDH, MB
-CRP, RF, ASO
SPECIFICATIONS |
|
Veterinary Biochemistry Analyzer |
|
Throughput |
Constant 180T/h (non-ISE) |
Theory |
Absorption spectrometry, transmission turbidimetry |
Methodology |
End-point、Fixed-time、Kinetic、Ion selective electrode |
Test according set up, random access |
|
Reaction plate |
Contain 81 reaction cuvettes |
Maximum reaction time |
15 minutes |
Minimum reaction volume |
150 ul |
Material of cuvettes |
Special UV-transparent optical glass |
Cuvettes use |
Automatically cuvettes washing |
Optical Unit |
|
Light Source |
Halogen-tungsten lamp |
Photometer |
Reversed optics |
Wavelength |
340nm-900nm wavelength option |
Absorption |
0-4 Abs |
Reagent Unit |
|
Reagent plate |
60 reagent positions,support for single and double reagent, reagent completely open |
Reagent storage temperature |
2~10℃ |
Reagent bottle |
15ml, 20ml, 30ml, 50ml or 70ml,dead volume is less than 1.0 ml |
Reagent volume |
20-500 ul, step by 1.0 ul |
Reagent/Sample Probe |
Automatically checked;Patent technology liquid level detection;Collision protection probe cleaning:Automatic washing for both inside and outside. |
Sample Unit |
|
Sample plate |
60 sample positions |
Sample type |
Human (animal) serum、plasma、urine etc. |
Samples cup |
Support one-off micro cups |
Sample volume |
2-100 ul, step by 0.1 ul |
Sample/reagent probe |
Automatically checked;Patent technology liquid level detection;Collision protection probe cleaning:Automatic washing for both inside and outside. |
Temperature control Unit |
|
Heating method |
Direct thermal solid thermostat closed |
Temperature equilibrium time |
<3 minutes |
Reaction temperature |
37±0.1℃ |
Temperature fluctuation |
±0.1℃ |
Operation Unit |
|
Software |
WindowsXP、2000 |
Database |
SQL Server |
Data Storage |
Automatic store & copy; No time and volume limit; Lifetime preservation |
Print format |
Can be set up, offer several standard format |
Calibration mode |
Linear curve type: Single-point linear, two-point linear and multi-point linear none linear; Curve type: Logistic-Log 4P,Logistic-Log5p,Spline,index automatically use best calibration curve and standard |
Quality control methods: |
Real time, within day & between day |
Quality control diagram |
L-J、cumulative、twin plot |
Status display |
Real time monitoring of reagent residual volume。Running status of reagent tray, samples tray |
Adjustment |
Dynamic and real time display of running status; Intelligent carry over settings to adjust test sequence; original reaction data record; Detail profile of alert messages |
Setting |
Operation permission setting, Item setting, cross pollution setting, cuvette blank setting |
Others |
|
Water consumption: <6L/h |
Power Supply: AC 220V±22V 50Hz±1Hz |
Frequency: Maximum 600VA |
Dimension: 782mm×581mm×545mm |
Net Weight: 60kg |
Accessory: Software |